Electro Surgery Unit

SURTRON 80 // 120 // 160

SURTRON® is a high frequency electrosurgical equipment which is suitable to light and medium surgery.

SURTRON®, through its performance, allows pure CUT, cut-coagulation BLEND, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG and, with a special adapter, BIPOLAR coagulation. The digital reading of delivered power and the monitoring of operative function by microcontroller guarantees the absolute reliability of working conditions.

SURTRON® allows a highly professional surgery thanks to the user-friendly and safety solutions normally used. The connection of neutral electrode is constantly monitored. Safety control of patient / plate contact using split neutral electrode. The possibility to control by the handle the monopoly output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.

80

Reference Code: 10100.101

Maximum Output Power CUT: 80 W - 250 Ω

Maximum Output Power BLEND: 60 W - 200 Ω

Maximum Output Power FORCED COAG: 50 W - 150 Ω

Maximum Output Power SOFT COAG: 40 W - 100 Ω

Maximum Output Power BIPOLAR: 30 W - 100 Ω

Working Frequency: 600 KHz

Patient Circuit: F

Selectable Input Voltage: 115 - 230 Vac

Mains Frequency: 50 - 60 Hz

Electrical Input Power: 230 VA

Size W x H x D mm: 254 x 104 x 288

Weight: 5Kgs

120

Reference Code: 10100.201

Maximum Output Power CUT: 120 W - 250 Ω

Maximum Output Power BLEND: 90 W - 200 Ω

Maximum Output Power FORCED COAG: 80 W - 150 Ω

Maximum Output Power SOFT COAG: 60 W - 100 Ω

Maximum Output Power BIPOLAR: 40 W - 100 Ω

Working Frequency: 600 KHz

Patient Circuit: F

Selectable Input Voltage: 115 - 230 Vac

Mains Frequency: 50 - 60 Hz

Electrical Input Power: 230 VA

Size W x H x D mm: 254 x 104 x 288

Weight: 5Kgs

160

Reference Code: 10100.301

Maximum Output Power CUT: 160 W - 250 Ω

Maximum Output Power BLEND: 120 W - 200 Ω

Maximum Output Power FORCED COAG: 100 W - 150 Ω

Maximum Output Power SOFT COAG: 80 W - 100 Ω

Maximum Output Power BIPOLAR: 60 W - 100 Ω

Working Frequency: 600 KHz

Patient Circuit: F

Selectable Input Voltage: 115 - 230 Vac

Mains Frequency: 50 - 60 Hz

Electrical Input Power: 230 VA

Size W x H x D mm: 254 x 104 x 288

Weight: 5Kgs

80

120

160

Reference Code

Maximum Output Power CUT

Maximum Output Power BLEND

Maximum Output Power FORCED COAG

Maximum Output Power SOFT COAG

Maximum Output Power BIPOLAR

Working Frequency

Patient Circuit

Selectable Input Voltage

Mains Frequency

Electrical Input Power

Size W x H x D mm

Weight

10100.101

80 W - 250 Ω

60 W - 200 Ω

50 W - 150 Ω

40 W - 100 Ω

30 W - 100 Ω

600 KHz

F

115 - 230 Vac

50 - 60 Hz

230 VA

254 x 104 x 288

5Kgs

10100.201

120 W - 250 Ω

90 W - 200 Ω

80 W - 150 Ω

60 W - 100 Ω

40 W - 100 Ω

600 KHz

F

115 - 230 Vac

50 - 60 Hz

230 VA

254 x 104 x 288

5Kgs

10100.301

160 W - 250 Ω

120 W - 200 Ω

100 W - 150 Ω

80 W - 100 Ω

60 W - 100 Ω

600 KHz

F

115 - 230 Vac

50 - 60 Hz

230 VA

254 x 104 x 288

5Kgs

Possibility of use for minimum evasive surgery

Possibility of use of bipolar forceps

Possibility of use of split neutral electrodes

Planning of the working conditions

Memorisation of the last used settings

Activation of the monopoly power with pedal and / or handle

Digital regulation and indication of the output power

Adjustment of the volume

Electro Surgery Unit_1

Standard Composition:

Reusable handle with switches F4243

Assorted electrodes (10 pcs) 00500.00

Cable for neutral electrode 00404.08

Steel neutral electrode 5365

Foot switch 00304.00

Power supply cable 2 mt 00100.03

Electro Surgery Unit_2

Safety:

EN60601-1

EN60601-1-2

EN60601-2-2

Electrical Class: I CF

MDD 93 / 42 / EC Class: II b

Electro Surgery Unit_3

Controls:

Patient / Plate circuit monitoring

Output power monitoring

Self check control